Quality control test of fixed x-ray units at Bir hospital and National trauma center

Authors

  • Kanchan Prasad Adhikari National Academy of Medical Sciences, Bir Hospital
  • Kanchan Sarga Nyaichyai National Academy of Medical Sciences, Bir Hospital
  • Birendra Bahadur Chand Bheri zonal hospital, Nepalgunj
  • Hanson Khadka National Academy of Medical Sciences, Bir Hospital
  • Saroj Sharma National Academy of Medical Sciences, Bir Hospital
  • Tara Khanal National Academy of Medical Sciences, Bir Hospital

DOI:

https://doi.org/10.3126/sw.v15i15.45648

Keywords:

Acceptance test; Image; Quality control; Radiation safety

Abstract

One of the important goals of the use of X-rays is to diagnose a disease and its extent and minimizing the adverse effect. In diagnostic X-ray, there is an important issue concerning the quality of the image and the exposure of radiation to the patient. Regular practice of quality control in diagnostic X-ray is essential to provide good quality images, decrease hazard to patients, and enhance the appropriate diagnosis. Quality control tests were carried out at six fixed X-ray units at National Academy of Medical Sciences, Bir Hospital and National Trauma Center. Parameters like kVp accuracy, timer accuracy, Reproducibility of kV, time and dose, variable mA and radiation output Linearity with the acceptance limits were checked. Radiation leakage, spatial resolution, contrast resolution and field alignment tests were also carried out. Quality control tests of kVp dose, dose rate, HVL, mAs, beam alignment and leakage radiation shows mixed results. The measured values were within limit in four X-ray units. Until now we do not have any regulations in Nepal on acceptance testing of x-ray units after installation. The need for quality assurance of X-ray needs to be taken to avert detrimental effects to patients and staff. Institutes must introduce a mandatory system for acceptance tests of X-ray when installed and alongside regular quality control programs.

Downloads

Download data is not yet available.
Abstract
159
PDF
188

Downloads

Published

2022-06-14

How to Cite

Adhikari, K. P., Nyaichyai, K. S., Chand, B. B., Khadka, H., Sharma, S., & Khanal, T. (2022). Quality control test of fixed x-ray units at Bir hospital and National trauma center. Scientific World, 15(15), 65–69. https://doi.org/10.3126/sw.v15i15.45648

Issue

Section

Research Article