Relationship of the Mean Glandular Dose with Compressed Breast Thickness in Digital Mammography

Authors

DOI:

https://doi.org/10.3126/njr.v10i1.28539

Keywords:

Compressed Breast Thickness, Mammogram, Mean Glandular Dose

Abstract

Introduction: Mammography is one of the most commonly performed radiological investigations for evaluation of breast cancer. As it involves ionizing radiation, there remains a risk of radiation induced cancer. In this study, we evaluated compressed breast thickness (CBT) and mean glandular dose (MGD) during routine mammography.

Methods: This prospective study was performed in the Department of Radiology and Imaging of TUTH. Data of 500 consecutive patients who underwent mammography over a period of 4 months (June 2018 to September 2018) was collected. The age, CBT and MGD were recorded. Pearson correlation and paired-t tests were performed.

Results: Most of the patients belonged to 41-50 years age group. MGD was significantly higher in patients with increased CBT. The CBT and MGD was higher in MLO view compared to CC views. The mean value of total MGD for four views was 5.1±1.4 mGy. There was significant positive correlation (r= 0.517) between CBT and MGD with increase in MGD with increase in CBT.

Conclusions: The MGD and dose equivalent in our routine mammography is within the recommended limits. MGD increases with increasing CBT and vice-versa. Hence, decreasing the thickness of compressed breast, can decrease the amount of radiation absorbed by the glandular tissue of the breast.

 

Downloads

Download data is not yet available.
Abstract
275
pdf
360

Author Biographies

Samjhana Khadka, Institute of Medicine, TUTH, Maharajgunj, Nepal

Radiographer

Department of Radiology 

Anamika Jha, Institute of Medicine, TUTH, Maharajgunj, Nepal

Assistant Professor

Department of Radiology 

Ranjit Kumar Chaudhary, University of Texas Medical Branch, Galveston, Texas, USA

Neuroradiology fellow

Department of Radiology

Shanta Lall Shrestha, Institute of Medicine, TUTH, Kathmandu, Nepal

Professor

Department of Radiology

Downloads

Published

2020-12-01

How to Cite

Khadka, S., Jha, A., Chaudhary, R. K., & Shrestha, S. L. (2020). Relationship of the Mean Glandular Dose with Compressed Breast Thickness in Digital Mammography. Nepalese Journal of Radiology, 10(1), 11–15. https://doi.org/10.3126/njr.v10i1.28539

Issue

Section

Original Articles