A comparative study of intraocular pressure measurement by three tonometers in normal subjects

Authors

  • SK Prabhakar JSS Medical College & Hospital, MG Road, Mysore-570004, Karnataka,
  • BS Mahesh JSS Medical College & Hospital, MG Road, Mysore-570004, Karnataka,
  • M Shanthamallappa JSS Medical College & Hospital, MG Road, Mysore-570004, Karnataka,

DOI:

https://doi.org/10.3126/nepjoph.v5i2.8729

Keywords:

iCare tonometer, Keeler Pulsair tonometer, Goldmann applanation tonometer, Perkins tonometer, Intraocular pressure

Abstract

Introduction: Contact Goldmann applanation tonometry (GAT) is gold standard for measuring intraocular pressure; however its routine use is limited by its non-portability and the need for a Slit Lamp Microscope. The Portable Perkins tonometer is also considered gold standard because it is based on the same principles as the GAT. The iCare is a newly introduced, portable, non-contact tonometer (NCT) that measures intraocular pressure (IOP) using a thin metallic probe.

Objective: To evaluate reliability and accuracy of IOP measurements using iCare and Keeler Pulsair tonometers against Perkins tonometer.

Subjects and methods: A comparative, randomized, prospective clinical study conducted on 166 eyes of 83 (n=83) subjects in the age group 14 to 71 years. The pressures were first recorded by iCare and Pulsair and then by Perkins. The SPSS 11.00 version was used for analysis.

Results: Mean pressures and standard deviation (+/-SD) for iCare, Pulsair and Perkins were 14.62(+/- 2.47), 14.53(+/-3.36) and 13.06(+/-2.69) and the Standard Error of the Mean (SEM) was 0.27, 0.36 and 0.30 respectively. There was a good correlation between iCare and Perkins with statistically significant difference (r=0.610, p <0.05). Regression analysis was performed. Using the Bland-Altman analysis 95% Limits of Agreement (LoA) for iCare and Pulsair were determined as -6.1 to 2.9 and -4.5 to 7.5 respectively.

Conclusion: Although both tonometers overestimated the Perkins values, Pulsair showed a better agreement with Perkins tonometer than iCare tonometer.

Nepal J Ophthalmol 2013; 5(10): 201-206

DOI: http://dx.doi.org/10.3126/nepjoph.v5i2.8729

Downloads

Download data is not yet available.
Abstract
1142
PDF
920

Downloads

Published

2013-09-25

How to Cite

Prabhakar, S., Mahesh, B., & Shanthamallappa, M. (2013). A comparative study of intraocular pressure measurement by three tonometers in normal subjects. Nepalese Journal of Ophthalmology, 5(2), 201–206. https://doi.org/10.3126/nepjoph.v5i2.8729

Issue

Section

Original Articles