Intraocular lens tilt and decentration: A concern for contemporary IOL designs

Authors

  • Jit B Ale

DOI:

https://doi.org/10.3126/nepjoph.v3i1.4281

Keywords:

IOL misalignment, Intraocular Lens, modern IOL, consequences of misalignment

Abstract

Purpose: To review published studies reporting the posterior chamber intraocular lens tilt and decentration after surgically uneventful implantation. Potential influences of normally occurring misalignment of modern designs of IOL on the optical performances are discussed.

Materials and methods: Published theoretical and clinical studies in relation to primarily implanted posterior chamber intraocular lenses and reports relating to more recent development of intraocular lens technologies were reviewed.

Results: Capsulotomy type and integrity, ocular pathology, fixation position of the haptics are some of the important factors causing the misalignment. On an average, a 2-3 degrees tilt and a 0.2 -0.3 mm decentration are common, and which remain clinically unnoticed for any design of IOL. However, theoretical studies predict deterioration of retinal image quality particularly with customized wavefront correcting IOLs. More than a 10 degrees tilt and above 1 mm decentration are occasionally reported even with modern cataract surgery in about 10 % of pseudophakic population.

Conclusions: The rate and extent of the complication have lowered substantially concomitant with developments in surgical techniques and IOL designs. While emerging designs of modern IOLs offer improved quality of postoperative vision, optimum performance is vastly influenced by the position of the device in the eye. Therefore, additional precision in alignment of modern designs of IOL may be warranted.

Key words: IOL misalignment; Intraocular Lens; modern IOL; consequences of misalignment

DOI: 10.3126/nepjoph.v3i1.4281

Nepal J Ophthalmol 2011;3(5):68-77

Downloads

Download data is not yet available.
Abstract
1417
PDF
1285

Author Biography

Jit B Ale

a. Vision Cooperative Research Centre, Sydney, NSW, Australia
b. Institute for Eye Research, Sydney, NSW, Australia
c. School of Optometry and Vision Science, University of New South Wales, Sydney, Australia

Downloads

How to Cite

Ale, J. B. (2011). Intraocular lens tilt and decentration: A concern for contemporary IOL designs. Nepalese Journal of Ophthalmology, 3(1), 68–77. https://doi.org/10.3126/nepjoph.v3i1.4281

Issue

Section

Review Article