Comparison of I-Gel and classic Laryngeal Mask Airway in paediatric population: a parallel group study
DOI:
https://doi.org/10.3126/jsan.v3i2.15618Keywords:
Airway, I -gel, Laryngeal mask airway, supraglottic devicesAbstract
Background: I-gel, the novel Supraglottic airway device, is easier to insert, has improved stability after insertion with reduced tissue compression.The study aims to compare the efficacy of I-gel with classic Laryngeal Mask Airway in the paediatric population.
Methods: A randomised parallel group study was done in Department of Anesthesiology, Kanti Children Hospital, Kathmandu for a period of three months. I gel and classic LMA was compared based on leak airway pressure, time of insertion and ease of insertion.
Results: Age and weight are comparable among groups. Compared to cLMA, I-gel provides a better leak pressure seal (16.40+-3.42 vs. 23.11+- 6.17 cm of H2O, p 0.027), faster time of insertion (19.42+-4.40 vs. 29.84+-7.70 seconds, p-0.02) and similar ease of insertion (p-0.571).
Conclusions: I-gel compared to classic Laryngeal Mask Airway provides better resistance to leak airway pressure, faster time of insertion with comparable ease of insertion.
Downloads
Downloads
Published
How to Cite
Issue
Section
License
© Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.