Agreement of results between dry and wet chemistry system for common clinical chemistry parameters

Authors

  • Saroj Thapa Kathmandu University School of Medical Sciences https://orcid.org/0000-0003-1234-644X
  • Apeksha Niraula Tribhuvan University Teaching Hospital, Institute of Medicine, Maharajgunj, Kathmandu, Nepal
  • Prabodh Risal Kathmandu University School of Medical Sciences

DOI:

https://doi.org/10.3126/jpn.v13i2.55219

Keywords:

Dry chemistry, Wet chemistry, Biochemical parameters, Aggregement

Abstract

Background: The advancement of analytical chemistry has led to the use of automated systems in clinical laboratories, including liquid-based wet chemistry and film-based dry chemistry systems. Reflectance spectrophotometry-based dry chemistry systems offer a viable alternative to wet chemistry analyzers. This study aimed to compare the agreement of results between the dry chemistry system and the wet chemistry system.

Materials and Methods: Secondary data from external quality control samples were analyzed using the Vitros 350 and BA-400 analyzers, over a period of one year (from 1st August 2021 to 31st July 2022). A total of 57 samples and 12 biochemical parameters were considered. Statistical analysis, including paired t-tests, Spearman's correlation, Bland-Altman plots, and the Intraclass Correlation Coefficient were applied. p value <0.05 was considered to be statistically significant.

Results: The bland-Altman analysis demonstrated that the measurements from both methods fell within the 95% limits of agreement for most of the clinical chemistry parameters like glucose, urea, creatinine, and liver enzymes indicating overall agreement. ICC values indicated excellent reliability for 10 out of 12 parameters, with HDL-C showing moderate reliability and albumin demonstrating good reliability respectively.

Conclusions: The findings of the present study suggest a high level of agreement and correlation between the dry and wet chemistry systems for common biochemical parameters. However, it is important to consider the specific parameters and limitations of each system. These results have implications for laboratories and healthcare professionals in selecting the most suitable system for their needs and resources.

Downloads

Download data is not yet available.
Abstract
385
PDF
178

Author Biographies

Saroj Thapa, Kathmandu University School of Medical Sciences

Lecturer, Department of Clinical Biochemistry

Apeksha Niraula, Tribhuvan University Teaching Hospital, Institute of Medicine, Maharajgunj, Kathmandu, Nepal

Assistant Professor, Department of Clinical Biochemistry

Prabodh Risal, Kathmandu University School of Medical Sciences

Professor, Department of Clinical Biochemistry

Downloads

Published

2023-12-31

How to Cite

Thapa, S., Niraula, A., & Risal, P. (2023). Agreement of results between dry and wet chemistry system for common clinical chemistry parameters. Journal of Pathology of Nepal, 13(2), 2059–2065. https://doi.org/10.3126/jpn.v13i2.55219

Issue

Section

Original Articles

Similar Articles

<< < 1 2 3 4 5 6 7 8 > >> 

You may also start an advanced similarity search for this article.